Up To 225 Watt Regulated DC Output High Voltage
Power Supplies for Laboratory or Systems Applications
- Up To 50 kV Output
Reversible Polarity
Short Circuit and Arc Protected
Custom and OEM Designs
Remote Analog Programming
Low Ripple and Noise
Remote Monitoring
CE Certified
General | Output | Input | Performance | Features | Mechanical | Accessories | Options | Custom ModelsGeneral
The Series 210 is a family of precision regulated linear power supplies with output voltages up to 50kV. The units are fully enclosed and can be operated as bench top instruments or mounted into a 19'' rack. These stable, low noise high voltage power supplies feature reversible polarity, remote analog programming and monitoring, front panel voltage and current metering, and calibrated direct reading front panel voltage controls. All units have arc and short circuit protection for safe, reliable operation.
The output high voltage of the Series 210 can be remotely programmed or controlled via the precision front panel direct reading controls. All models can be adjusted over their full output voltage range using an analog programming 0 to -5V signal or using an external potentiometer.
Remote analog monitoring of the high voltage output is standard on all Series 210 power supplies. Signals proportional to the output voltage and output current are provided at the rear panel I/O connector. A logic output signal indicating high voltage polarity is also standard.
Output
Voltage and Current:
See Chart below
Polarity:
For 1kV through 5kV models polarity reversal is achieved by a screwdriver type switch located on the rear panel of the unit.
For 10kV through 50kV models polarity reversal is achieved by the reversal of an internal connector. The selected polarity is displayed on a front panel LED indicator.
MODEL VOLTAGE CURRENT RIPPLE
(pk-pk)210-01R 0 to 1kV 0 to 225mA 50mV 210-01.5R 0 to 1.5kV 0 to130mA 100mV 210-02R 0 to 2kV 0 to100mA 100mV 210-03R 0 to 3kV 0 to 75mA 100mV 210-05R 0 to 5kV 0 to 40mA 200mV 210-10R 0 to 10kV 0 to 15mA 500mV 210-20R 0 to 20kV 0 to 7mA 1V 210-30R 0 to 30kV 0 to 4.5mA 1.5V 210-50R 0 to 50kV 0 to 2.5mA 5V All units have reversible polarity. Ripple is measured peak to peak at maximum output.
Input
Power:
115 Vac ±10% @ 5 Amperes (6.25A for models up to 5kV), 50-60 Hz.
230 Vac ±10% @ 3.0 Amperes, 50-60Hz.
Performance
Line Regulation:
±0.001% of maximum for +10% input line change.
Load Regulation:
±0.005% of maximum for 0 to maximum output current change.
Ripple:
See Chart above
Temperature Coefficient (0 to 50°C):
50ppm of maximum per °C.
Storage Temperature:
-40°C to +85°C
Stability (after ½ hr warm-up):
0.01% per hour; 0.02% per 8 hours.
Features
Front Panel Meter:
Front panel analog meter, switch selectable for reading output voltage and current. The meter accuracy is ±2% of full scale.
Front Panel Controls:
Calibrated front panel direct reading multi-turn precision potentiometer and switches. Accuracy is ±(0.25% of setting + 0.05% of maximum) for models up to and including 30kV; ±(0.5% of setting + 0.05% of maximum) for 50kV model.
Resolution:
0.2V for models up to and including 30kV; 20V for 50kV model.
Remote Programming:
0 to -5 Volt dc analog input signal proportional to 0 to maximum rated output. Accuracy is ±(0.25% of setting + 0.05% of maximum) for models up to and including 30kV; ±(0.5% of setting + 0.05% of maximum) for 50kV model. The programming input impedance is 5 kilohms.
Analog Output Voltage Monitor:
0 to +5 Volts proportional to 0 to maximum output high voltage. Accuracy is ±(0.25% of reading + 0.25% of maximum). The monitor output impedance is 50 kilohms.
Analog Output Current Monitor:
0 to +5 Volts proportional to 0 to maximum output current, except as indicated. Accuracy is ±(0.5% of reading + 0.25% of maximum). The monitor output impedance is 50 kilohms.
Current Limit:
Automatic current limiting occurs at approximately 110% of maximum rated output current at the maximum rated output voltage. The allowable maximum output current at any set voltage must be derated linearly from 100% at maximum voltage output down to 30% of maximum output current at 0 output voltage. Supply is self-restoring upon removal of cause of current limit condition.
Protection:
Arc and short circuit, self restoring.
Mechanical
Size:
All units are 19 inch (483mm) wide standard rack mount and 5¼ inches (133mm) high. Models up through 5kV are 11 inches (279mm) deep. All units above 5kV are 16 inches (406mm) deep.
Weight:
See Chart below
High Voltage Connector:
See Chart below
Note: Except where noted below, preassembled HV cable must be purchased separately.
MODEL OUTPUT MATING lbs (kg) 210-01R JAC PAE 34 (15) 210-01.5R JAC PAE 34 (15) 210-02R JAC PAE 34 (15) 210-03R JAC PAE 34 (15) 210-05R JAC PAE 34 (15) 210-10R JJA 405787 34 (15) 210-20R JJA 405787 37 (17) 210-30R JJA 405787 39 (18) 210-50R JJB 405787 46 (21) The mating high voltage connector is provided with each unit. Note: Bertan P/N PAE is equivalent to Kings SHV type 1705-14.
Power Input Connector:
A captive 3-wire line cord and NEMA plug is included.
Low Voltage I/O Connector:
The PROGRAMMING/MONITOR connector P/N JKB provides all remote control and monitor functions. The mating connector P/N PKB is provided.
Cooling:
Internal fan.Custom Models
The Series 210 can be economically and quickly modified to satisfy custom applications. Other output voltage and/or current ratings, custom control features, digital programming, or special mechanical constraints are some of the varied requirements which can be satisfied. Contact Bertan Sales Engineering for a responsive review of your application.
Adobe Acrobat Documents
The following documents are available in Adobe Acrobat (PDF) format. The Adobe Acrobat Reader allows you to view, navigate, and print PDF files across all major computor platforms. To get a free reader click on the "Get Acrobat Reader" icon. To view a document click on the title below. To download a document onto your computer hold down the SHIFT key and click on the title (Netscape) or right-click on the title and select "Save Target As ..." (Internet Explorer). Instructions may vary on some browsers and versions.